ABI Procise Solenoid Mounting Block parts

From: Chin, David T. (ChinD@missouri.edu)
Date: Mon Jan 03 2000 - 16:11:32 EST


        On the Procise valve block contain valve #45 to the flask, my
solenoid mounting block (the thing that holds the solenoid to the valve
block manifold) is pitted and causing the membrane to rupture prematurely.
The solenoid mounting block is held in place with 2 screws and two pins
instead of the 4 screws in the old design. Using the old style block causes
a misalignment and premature membrane rupture.
        Is there a source for the new style mounting blocks?

David T. Chin
Director, Protein Core Facility
Protein Chemistry and Expression
2-17 Agriculture Building
Univ. of Missouri - Columbia
Columbia, MO 65211-7170

SHIPPING AND ACTUAL LOCATION:
2-17 Agriculture Building (office)
2-31 Agriculture Building (Lab)

e-Mail: chind@missouri.edu
web: http://www.biotech.missouri.edu/pc
[mailto:chind@missouri.edu]
Phone: 573-882-2027
Fax: 573-882-7105



This archive was generated by hypermail 2b29 : Tue Jan 18 2000 - 11:43:38 EST