Re: HP1100/ buffer problems

From: Joe Gray (Joe.Gray@newcastle.ac.uk)
Date: Mon Feb 07 2000 - 05:14:09 EST


Dear Jeni

Is your system fitted with the thermostated column compartment?
In my 1100 all the columns are now fitted with in-line frits,
because I find that that even in 'normal' RP-HPLC use they tend
to block fairly quickly, necessitating a change about once a
week. Examination of the blocked frits often shows that they
have collected small pieces of 'rust'? I suspect that they may
be coming from the thermostated compartment - they look like
sprayed cast iron blocks to me. Running salt solutions through
these blocks is therefore bound to accelerate your problems - if
what I suspect is right that is - anyone from Agilent care to
comment?

Good Luck, JG.

**********************************