Re:Pepsyn

From: Radhika Krishnan (rkrishna@coh.org)
Date: Mon May 15 2000 - 12:58:58 EDT


Rebecca,
     I used to do synthesis on a VEGA machine which had no feedback
     monitoring. I think they now exist only in the graveyards of
     technical equipment . However the only way I could monitor a synthesis
     would be to do the trusted Ninhydrin test at every step, or every
     crucial step ( depending on the sequence). With fmoc synthesis however
     you might get false positives when testing a coupling, because the
     pyridine in test solution 2, might cleave the fmoc off of the resin.
     You will have to use a different reagent to test iminoacids like Pro
     ( isatin solution works). Hope this helps.
     Radhika.