Bob-
for calibration I have used the neg ultramark signal.
and I have seen methods using,
TFA solvent clusters under non-cluster busting source conditions (low temp/gas).
higher (0.1% NH4?) pH myoglobin solutions (infusion only usually)
and ivory dishwashing soap (watch the amount-contamination like ultramark)
http://www.jeol.com/ms/ivory.html
for source "tuning" I'd use a "standard" like your analyte.
Matt Sweeney
mattsweeney@earthlink.net
Mass Spec Consulting
Training/Operations/Consulting/Method Development
LC/MS Pharmacokinetics, Peptides, Proteins, Metabolism,
Maintenance Classes, Specialist in Finnigan Equipment and Software
-----Original Message-----
From: boblee <boblee@uic.edu>
To: Recipients of ABRF List <abrf@aecom.yu.edu>
Date: Wednesday, October 25, 2000 12:26 AM
Subject: Standards for negative ion mode
>Hi,
>
>Need to find a good and cheap compound (including the solvent system) for
>optimizing negative ion mode of LCQ. Thanks.
>
>
>Bob
>
>
>