Dong,
Lots of possibilities. Think "HPLC" and troubleshoot accordingly.
My first guess is that you went over-pressure, and the system shut itself
down automatically. Check the back-pressure on your system at 14%. If it's
high, try swapping in an old column, to begin to track down the plug. Other
places that 473s plug frequently include the sample injection valve and
associated sample loop.
I assume you're using a system that worked before, so the gradient command
set is probably OK. But you might check to make sure that you don't
inadvertently have a command for "0 ml/min flow rate" or "end analysis" at
the time for the start of the gradient.
Mark
-----Original Message-----
From: Zhidong Xie [mailto:zxie@ctl.net]
Sent: Wednesday, January 03, 2001 7:27 PM
To: Recipients of ABRF List
Subject: Help!!
Hello, everyone,
Happy New Year!
I have a problem with running 473A protein sequencer. The
problem is: when I ran a STD cycle, HPLC pumps started initial 12%B
for 14 mins, but it stopped pumping right before Start Gradient step,
i.e Gradient won't start. Did anybody come across similar problem
before?
Any suggestion is welcomed. Thank you very much!
Regards
Dong