Re: PE Newguard cartridges for 120A

Gary Hathaway (hathaway@cco.caltech.edu)
Sat, 14 Mar 1998 00:32:54 -0700

At 8:13 AM -0700 3/13/98, Mark Garfield wrote:
>Dear ABRF,
>
>I recently received some Newguard cartridges (ABI cat. no. 0711-0221)
>for the 120A HPLC systems on our 477 protein sequencers. All these new
>cartridges have been leaking, which makes me think that somehow these
>are different sizes than the old ones. They also come in different
>packaging....clear plastic as opposed to cardboard boxes.
>
>Has anyone else experienced this problem?
>Mark K. Garfield

Mark,
Our field service engineer told me (rather unapologetically) the new guard
cartridges are of smaller O.D. His suggestion was that this could be fixed
if you replaced the end seals in the Brownlee holder. So far, I tried one
and it leaked so I'm using up the rest of my stock of old cartridges before
disassembling the holder. Best of luck.

--------------------------------------------------