Re: HPLC: 1 mm columns

John D. Shannon (jds1c@avery.med.virginia.edu)
Tue, 12 May 1998 16:33:35 -0400 (EDT)

According to Mark Lively:
>
> ABRF colleagues,
>
> Similarly, can the ABI Model 120 HPLC system be used for
> 1 mm columns? I would assume it would be necessary to
> change or replace the relavtively large volume dynamic
> mixers on each.
> Thanks in advance for your input.
> -Mark
>
> Mark O. Lively, Ph.D.
> Professor of Biochemistry
> Wake Forest University School of Medicine
> Medical Center Blvd.
> Winston-Salem, North Carolina 27157
> Phone: 336-716-2969
> Fax: 336-716-7200
> email: mlively@wfubmc.edu
>
>
I have used 1 mm columns on an ABI 120 for the last year, using
a 75 ul mixer and low volume flow cell, both from ABI.

To mount the mixer (probably made for the 140), I used a piece
of angle aluminum, and bolted on the motor from the old mixer
to save changing the electrical connections (a suggestion from
a service engineer).

Initially the baseline was bad but now I can run at 0.02 AUFS.

-- 
John Shannon