Re: HPLC: 1 mm columns
David A. Schooley (schooley@unr.edu)
Tue, 12 May 1998 16:45:20 -0700
Mark-
You better check your specs on the instrument, but I believe that
the detector volume is 3 microliter or greater. This is NOT good for 1 mm
columns. We are doing 1 mm chromatography with an HP 1100 detector with a
1 ul flow cell with good peak shape. The Michrom LC's have flow cells with
50 nanoliter or lower cell volumes. On a Michrom you have the peak
collected in a tube while it is still mixing in the cell of an ABI 120.
David
>ABRF colleagues,
>
> I would like to know if it is possible to use an Applied
>Biosystems Model 140B Solvent Delivery system to run 1 mm
>i.d. columns. The detector is a Model 785A absorbance