Re: DNA- genescan

Deb Grove (dsg4@psu.edu)
Fri, 10 Sep 1999 18:54:26 -0400

1) You could reinstall the matrix.

2) I had trouble once when the CCD offset was not 250 as the factory specs
usually are. My machine had shifted to 211. The specialist brought up all
kinds of other things to try but when the field technician checked it out,
this turned out to be the answer. Our other machine is at 250.

Deb G.

Deborah S. Grove, Ph.D.
Manager, Nucleic Acid Facility
210 Wartik Laboratory
Life Science Consortium
The Pennsylvania State University
University Park, PA 16802
Tel: (814)865-3332

http://www.biotec.psu.edu/lsc/stf/naf/naf.html