Re: AAA AccQT. question?

Pawel Mak (makp@mol.uj.edu.pl)
Mon, 18 Oct 1999 08:59:13 +0200

James,
Do you keep the times between injections (or runs of a gradients) equal ?
>From my experience it is very important factor for reproductibility of the
retention times, when I do amino acid analysis the column works non stop,
gradient after gradient with no pauses between them (even if the sample is
not injected).
The second possibility I wonder is the unstable temperature housing of the
column..., evaporating of acetonitrile from solvents...???
Cheers,
Pawel