MS Calibration question?

DOUGLAS LAMONT (dougie01@lamontjka.freeserve.co.uk)
Wed, 10 Nov 1999 00:24:08 -0000

This is a multi-part message in MIME format.

------=_NextPart_000_00B3_01BF2B11.E771DA60
Content-Type: text/plain;
charset="iso-8859-1"
Content-Transfer-Encoding: quoted-printable

Hi all,

I work with a single quad system (Sciex API100). I usually calibrate =
using
PPGs then use apomyoglobin (mw=3D16951.5 Da) to check mass accuracy. =
Recently
I've had problems with the accuracy side. The PPG calibration still =
works
fine and is within spec. i.e. linear over the mass range, but the apo. =
mass
is out using the hypermass algorithm. I usually get 16953 (0.009%) for =
apo.=20
which is OK but I now consistently get 16954(0.015%). After running the =
apo.
I recheck the PPGs and again they are in spec.=20

any help would be appreciated

Thanks in advance

Dougie Lamont

PPL Therapeutics

------=_NextPart_000_00B3_01BF2B11.E771DA60
Content-Type: text/html;
charset="iso-8859-1"
Content-Transfer-Encoding: quoted-printable

<!DOCTYPE HTML PUBLIC "-//W3C//DTD W3 HTML//EN">

Hi all,

I work with a single quad system (Sciex API100). I = usually=20 calibrate using
PPGs then use apomyoglobin (mw=3D16951.5 Da) to check = mass=20 accuracy. Recently
I've had problems with the accuracy side. The PPG=20 calibration still works
fine and is within spec.  i.e. linear = over the=20 mass range, but the apo. mass
is out using the hypermass algorithm. I = usually=20 get 16953 (0.009%) for apo.
which is OK but I now consistently get 16954(0.015%). After running = the=20 apo.
I recheck the PPGs and again they are in spec.

any help = would be=20 appreciated

Thanks in advance

Dougie Lamont